Design of a Skipper CCD Focal Plane for the SOAR Integral Field Spectrograph (2210.03665v1)
Abstract: We present the development of a Skipper Charge-Coupled Device (CCD) focal plane prototype for the SOAR Telescope Integral Field Spectrograph (SIFS). This mosaic focal plane consists of four 6k $\times$ 1k, 15 $\mu$m pixel Skipper CCDs mounted inside a vacuum dewar. We describe the process of packaging the CCDs so that they can be easily tested, transported, and installed in a mosaic focal plane. We characterize the performance of $\sim 650 \mu$m thick, fully-depleted engineering-grade Skipper CCDs in preparation for performing similar characterization tests on science-grade Skipper CCDs which will be thinned to 250$\mu$m and backside processed with an antireflective coating. We achieve a single-sample readout noise of $4.5 e{-} rms/pix$ for the best performing amplifiers and sub-electron resolution (photon counting capabilities) with readout noise $\sigma \sim 0.16 e{-} rms/pix$ from 800 measurements of the charge in each pixel. We describe the design and construction of the Skipper CCD focal plane and provide details about the synchronized readout electronics system that will be implemented to simultaneously read 16 amplifiers from the four Skipper CCDs (4-amplifiers per detector). Finally, we outline future plans for laboratory testing, installation, commissioning, and science verification of our Skipper CCD focal plane.
Sponsor
Paper Prompts
Sign up for free to create and run prompts on this paper using GPT-5.
Top Community Prompts
Collections
Sign up for free to add this paper to one or more collections.