Papers
Topics
Authors
Recent
2000 character limit reached

Calibration of a two-phase xenon time projection chamber with a $^{37}$Ar source

Published 24 May 2017 in physics.ins-det | (1705.08958v3)

Abstract: We calibrate a two-phase xenon detector at 0.27 keV in the charge channel and at 2.8 keV in both the light and charge channels using a ${37}$Ar source that is directly released into the detector. We map the light and charge yields as a function of electric drift field. For the 2.8 keV peak, we calculate the Thomas-Imel box parameter for recombination and determine its dependence on drift field. For the same peak, we achieve an energy resolution, $E_{\sigma}/E_{mean}$, between 9.8% and 10.8% for 0.1 kV/cm to 2 kV/cm electric drift fields.

Citations (34)

Summary

We haven't generated a summary for this paper yet.

Whiteboard

Paper to Video (Beta)

Open Problems

We haven't generated a list of open problems mentioned in this paper yet.

Continue Learning

We haven't generated follow-up questions for this paper yet.

Collections

Sign up for free to add this paper to one or more collections.