Accurate Patient Alignment without Unnecessary Imaging Dose via Synthesizing Patient-specific 3D CT Images from 2D kV Images (2405.19338v1)

Published 1 Apr 2024 in eess.SP, cs.AI, and cs.CV

Abstract: In radiotherapy, 2D orthogonally projected kV images are used for patient alignment when 3D-on-board imaging(OBI) unavailable. But tumor visibility is constrained due to the projection of patient's anatomy onto a 2D plane, potentially leading to substantial setup errors. In treatment room with 3D-OBI such as cone beam CT(CBCT), the field of view(FOV) of CBCT is limited with unnecessarily high imaging dose, thus unfavorable for pediatric patients. A solution to this dilemma is to reconstruct 3D CT from kV images obtained at the treatment position. Here, we propose a dual-models framework built with hierarchical ViT blocks. Unlike a proof-of-concept approach, our framework considers kV images as the solo input and can synthesize accurate, full-size 3D CT in real time(within milliseconds). We demonstrate the feasibility of the proposed approach on 10 patients with head and neck (H&N) cancer using image quality(MAE: <45HU), dosimetrical accuracy(Gamma passing rate (2%/2mm/10%)>97%) and patient position uncertainty(shift error: <0.4mm). The proposed framework can generate accurate 3D CT faithfully mirroring real-time patient position, thus significantly improving patient setup accuracy, keeping imaging dose minimum, and maintaining treatment veracity.

Summary

We haven't generated a summary for this paper yet.

X Twitter Logo Streamline Icon: https://streamlinehq.com