Resolve the discrepancy between alternative strontium bridge isotopes
Resolve the approximately 3% discrepancy between the Sr dose factors obtained using 86Sr and 88Sr as alternative bridge isotopes for linking the 9 ns and 20 ns phases in multiphase ToF-SIMS acquisition.
References
As an internal check (S6.4), 88Sr itself is also below 3 corrected counts per pulse at 20 ns and can serve as the bridge; the dose factor it gives differs from the 86Sr value by about 3%, and the Sr ratios computed with it lie 2–3% lower (86Sr/88Sr +1.3%, 87Sr/88Sr +0.5%, 84Sr/88Sr +4.6%). The disagreement is reported, not resolved; it does not affect the Ti ratios.
— Quantitative Interleaved-Pulse Acquisition Extends the Range of Isotope-Ratio Measurements in ToF-SIMS
(2609.37413 - Ievlev, 29 Sep 2026) in Supporting Information, Section S5.2, p. 6; see also main text, p. 10