Determine the source of the short-phase acquisition deficit
Determine the cause of the approximately 40% reduction in 18O counts and poorer crater-averaged precision observed when the short pulse phase is acquired last rather than first in interleaved-pulse ToF-SIMS acquisition.
References
Placing the short phase last (P3) gave about 40% fewer 18O counts in the short phase (824 versus 1375) and a poorer crater average (3.48%); the cause was not identified, and the short phase is acquired first throughout.
— Quantitative Interleaved-Pulse Acquisition Extends the Range of Isotope-Ratio Measurements in ToF-SIMS
(2609.37413 - Ievlev, 29 Sep 2026) in Supporting Information, Section S3, p. 4