Cause of residual organ-at-risk dose differences
Identify whether the residual organ-at-risk dose offset between uncertainty-guided digital twin online adaptive proton therapy plans and physician-approved offline replans is attributable to reuse of the offline replan objective template, use of patient-specific robust margins, or another component of the planning framework.
References
The online plans reused the objective template of the offline replan without further tuning and were optimized with the patient-specific margins of Table 2 as their setup uncertainty; which of these accounts for the OAR offset cannot be separated with the present data, and the offset kept eight of ten online plans within ±2% of the baseline score for either starting point.
— An Uncertainty-Guided Digital Twin Framework for Online Adaptive Proton Therapy in Head and Neck Cancer: A Feasibility Study