Metal artefact reduction sequences for a piezoelectric bone conduction implant using a realistic head phantom in MRI
Abstract: Industry standards require medical device manufacturers to perform implant-induced artefact testing in phantoms at a pre-clinical stage to define the extent of artefacts that can be expected during MRI. Once a device is commercially available, studies on volunteers, cadavers or patients are performed to investigate implant-induced artefacts and artefact reduction methods more in-depth. This study describes the design and evaluation of a realistic head phantom for pre-clinical implant-induced artefact testing in a relevant environment. A case study is performed where a state-of-the-art piezoelectric bone conduction implant is used in the 1.5 T and 3 T MRI environments. Images were acquired using clinical and novel metal artefact reducing (MARS) sequences at both field strengths. Artefact width and length were measured in a healthy volunteer and compared with artefact sizes obtained in the phantom. Artefact sizes are reported that are similar in shape between the phantom and a volunteer, yet with dimensions differing up to 20% between both. When the implant magnet is removed, the artefact size can be reduced below a diameter of 5 cm, whilst the presence of an implant magnet and splint creates higher artefacts up to 20 cm in diameter. Pulse sequences have been altered to reduce the scan time up to 7 minutes, while preserving the image quality. These results show that the anthropomorphic phantom can be used at a preclinical stage to provide clinically relevant images, illustrating the impact of the artefact on important brain structures.
Paper Prompts
Sign up for free to create and run prompts on this paper using GPT-5.
Top Community Prompts
Collections
Sign up for free to add this paper to one or more collections.