Papers
Topics
Authors
Recent
2000 character limit reached

An algorithm to calibrate and correct the response to unpolarized radiation of the X-ray polarimeter on board IXPE (2111.14867v1)

Published 29 Nov 2021 in astro-ph.IM and astro-ph.HE

Abstract: The Gas Pixel Detector is an X-ray polarimeter to fly on-board IXPE and other missions. To correctly measure the source polarization, the response of IXPE's GPDs to unpolarized radiation has to be calibrated and corrected. In this paper we describe the way such response is measured with laboratory sources and the algorithm to apply such correction to the observations of celestial sources. The latter allows to correct the response to polarization of single photons, therefore allowing great flexibility in all the subsequent analysis. Our correction approach is tested against both monochromatic and non-monochromatic laboratory sources and with simulations, finding that it correctly retrieves the polarization up to the statistical limits of the planned IXPE observations.