A Proton Computed Tomography Demonstrator for Stopping Power Measurements
Abstract: Particle therapy is an established method to treat deep-seated tumours using accelerator-produced ion beams. For treatment planning, the precise knowledge of the relative stopping power (RSP) within the patient is vital. Conversion errors from x-ray computed tomography (CT) measurements to RSP introduce uncertainties in the applied dose distribution. Using a proton computed tomography (pCT) system to measure the SP directly could potentially increase the accuracy of treatment planning. A pCT demonstrator, consisting of double-sided silicon strip detectors (DSSD) as tracker and plastic scintillator slabs coupled to silicon photomultipliers (SiPM) as a range telescope, was developed. After a significant hardware upgrade of the range telescope, a 3D tomogram of an aluminium stair phantom was recorded at the MedAustron facility in Wiener Neustadt, Austria. In total, 80 projections with 6.5x105 primary events were acquired and used for the reconstruction of the RSP distribution in the phantom. After applying a straight-line approximation for the particle path inside the phantom, the most probable value (MPV) of the RSP distribution could be measured with an accuracy of 0.59%. The RSP resolution inside the phantom was only 9.3% due to a limited amount of projections and measured events per projection.
Paper Prompts
Sign up for free to create and run prompts on this paper using GPT-5.
Top Community Prompts
Collections
Sign up for free to add this paper to one or more collections.