2000 character limit reached
Active Optical Table Tilt Stabilization
Published 6 Mar 2020 in physics.ins-det | (2003.03404v1)
Abstract: We show that a simple modification to an optical table with pneumatic vibration isolation can be used to actively reduce the long term drift in the tilt of the table by nearly a factor of 1000. Without active stabilization, we measure a root-mean-square (RMS) tilt variation of \SI{270}{\upmu rad} over three days. The active stabilization can be used to limit the tilt to \SI{0.35}{\upmu rad} RMS over the same time period. This technique can be used to minimize drift in tilt-sensitive experiments.