An InGrid based Low Energy X-ray Detector
Abstract: An X-ray detector based on the combination of an integrated Micromegas stage with a pixel chip has been built in order to be installed at the CERN Axion Solar Telescope. Due to its high granularity and spatial resolution this detector allows for a topological background suppression along with a detection threshold below $1\,\text{keV}$. Tests at the CAST Detector Lab show the detector's ability to detect X-ray photons down to an energy as low as $277\,\text{eV}$. The first background data taken after the installation at the CAST experiment underline the detector's performance with an average background rate of $5\times10{-5}\,/\text{keV}/\text{cm}2/\text{s}$ between 2 and $10\,\text{keV}$ when using a lead shielding.