Sub-keV Electron Recoil Calibration for Cryogenic Detectors using a Novel X-ray Fluorescence Source (2505.17686v1)
Abstract: Calibration of cryogenic detectors with energy thresholds below 100 eV are crucial for light Dark Matter (DM) searches and reactor neutrino studies based on coherent elastic neutrino-nucleus scattering (CEvNS). This paper presents a novel calibration source based on X-ray fluorescence (XRF) of light elements. It uses a ${55}$Fe source to irradiate a two-staged target arrangement, emitting monoenergetic X-rays from 677 eV to 6.5 keV. We demonstrate the potential of this new XRF source to calibrate a 0.75 gram CaWO$_4$ crystal of the NUCLEUS and CRAB experiments. Additionally, we introduce CryoLab, an advanced analysis tool for cryogenic detector data, featuring robust methods for data processing, calibration, and high-level analysis, implemented in MATLAB and HDF5. We also present a phenomenological model for energy resolution, which incorporates statistical contributions, systematic effects, and baseline noise, enabling a novel approach to evaluating athermal phonon collection efficiency in detectors based on transition edge sensors (TES).
Collections
Sign up for free to add this paper to one or more collections.
Paper Prompts
Sign up for free to create and run prompts on this paper using GPT-5.