A Novel Low-Background Photomultiplier Tube Developed for Xenon Based Detectors (2412.10830v2)
Abstract: Photomultiplier tubes (PMTs) are essential in xenon detectors like PandaX, LZ, and XENON experiments for dark matter searches and neutrino properties measurement. To minimize PMT-induced backgrounds, stringent requirements on PMT radioactivity are crucial. A novel 2-inch low-background R12699 PMT has been developed through a collaboration between the PandaX team and Hamamatsu Photonics K.K. corporation. Radioactivity measurements conducted with a high-purity germanium detector show levels of approximately 0.08 mBq/PMT for $\rm{60}Co$ and 0.06~mBq/PMT for the $\rm{238}U$ late chain, achieving a 15-fold reduction compared to R11410 PMT used in PandaX-4T. The radon emanation rate is below 3.2 $\rm \mu$Bq/PMT (@90\% confidence level), while the surface $\rm{210}Po$ activity is less than 18.4 $\mu$Bq/cm$2$. The electrical performance of these PMTs at cryogenic temperature was evaluated. With an optimized readout base, the gain was enhanced by 30\%, achieving an average gain of $4.23 \times 106$ at -1000~V and -100~${\circ}$C. The dark count rate averaged 2.5~Hz per channel. Compactness, low radioactivity, and robust electrical performance in the cryogenic temperature make the R12699 PMT ideal for next-generation liquid xenon detectors and other rare event searches.