Time-efficient filtering of polarimetric data by checking physical realizability of experimental Mueller matrices (2402.18555v1)
Abstract: Imaging Mueller polarimetry has already proved its potential for metrology, remote sensing and biomedicine. The real-time applications of this modality require both video rate image acquisition and fast data post-processing algorithms. First, one must check the physical realizability of the experimental Mueller matrices in order to filter out non-physical data, i.e. to test the positive semi-definiteness of the 4x4 Hermitian coherency matrix calculated from the elements of the corresponding Mueller matrix pixel-wise. For this purpose, we compared the execution time for the calculations of i) eigenvalues, ii) Cholesky decomposition, iii) Sylvester's criterion, and iv) coefficients of the characteristic polynomial of the Hermitian coherency matrix using two different approaches, all calculated for the experimental Mueller matrix images (600 pixels x 700 pixels) of mouse uterine cervix. The calculations were performed using C++ and Julia programming languages. Our results showed the superiority of the algorithm iv), in particular, the version based on the simplification via Pauli matrices, in terms of execution time for our dataset, over other algorithms. The sequential implementation of the latter algorithm on a single core already satisfies the requirements of real-time polarimetric imaging in various domains. This can be further amplified by the proposed parallelization (for example, we achieve a 5-fold speed up on 6 cores).
- T. Novikova and P. Bulkin. Inverse problem of Mueller polarimetry for metrological applications. J. Inverse Ill-Posed Probl., 29:759–774, 2021.
- Mueller matrix imaging ellipsometry for nanostructure metrology. Opt. Express, 23(13):17316–17329, 2015.
- Overlay measurements by Mueller polarimetry in back focal plane. J. Micro/Nanolith. MEMS MOEMS, 10(3):033017, 2011.
- Virtues of Mueller matrix imaging for underwater target detection. Appl. Opt., 38:6431–6438, 1999.
- Polarization fingerprint for microalgae classification. Opt. Lasers Eng., 166:107567, 2023.
- Mueller matrix imaging for collagen scoring in mice model of pregnancy. Sci. Rep., 11(1):15621, 2021.
- Mueller polarimetry of brain tissues. In Polarized light in Biomedical Imaging and Sensing, page 205–230. J. C. Ramella-Roman and T. Novikova (Eds), Springer, Sham, 2022.
- Tissue polarimetry. In Photonics: Biomedical Photonics Spectroscopy, and Microscopy. D. L. Andrews (Ed.), Wiley & Sons, Inc., 2015.
- Use of Mueller polarimetric imaging for the staging of human colon cancer. Proc. SPIE, Optical Biopsy IX, 7895:65–72, 2011.
- Backscattering Mueller matrix polarimetry on whole brain specimens shows promise for minimally invasive mapping of microstructural orientation features. Front. Photon., 3, 2022.
- Assessment of the impact of nanowarming on microstructure of cryopreserved fibroblast-containing 3D tissue models using Mueller polarimetry. Photonics, 10(10):1129, 2023.
- Mueller matrix measurement of electrospun fiber scaffolds for tissue engineering. Polymers, 11:2062, 2019.
- Digital histology with Mueller polarimetry and FastDBSCAN. Appl. Opt., 61(32):9616–9624, 2022.
- Mueller polarimetric microscopic images analysis based classification of breast cancer cells. Opt. Commun., 475:126194, 2020.
- 3D Mueller matrix mapping of layered distributions of depolarisation degree for analysis of prostate adenoma and carcinoma diffuse tissues. Sci. Rep., 11:5162, 2021.
- Combined Stokes vector and Mueller matrix polarimetry for materials characterization. Anal. Chem., 85(15):7613–7619, 2013.
- O. Arteaga and B. Kahr. Mueller matrix polarimetry of bianisotropic materials. J. Opt. Soc. Am. B, 36(8):F72–F83, Aug 2019.
- Applications of Mueller matrix polarimetry to biological and agricultural diagnostics: A review. Appl. Sci., 12(10):5258, 2022.
- D. H. Goldstein. Polarized light. (3rd ed. CRC Press), 2010.
- R. M. A. Azzam. Propagation of partially polarized light through anisotropic media with or without depolarization: A differential 4×\times×4 matrix calculus. J. Opt. Soc. Am., 68(12):1756 – 1767, 1978.
- J. J. Gil-Perez and R. Ossikovski. Polarized Light and the Mueller Matrix Approach. (2nd Ed. CRC Press), Boca Raton, FL, USA, 2022.
- Interpretation of Mueller matrices based on polar decomposition. J. Opt. Soc. Am.A, 13:1106–1113, 1996.
- S. R. Cloude. Group theory and polarisation algebra. Optik, 75:26–36, 1986.
- Analysis of tissue microstructure with Mueller microscopy: logarithmic decomposition and Monte Carlo modeling. J. Biomed. Opt., 25(1):015002, 2020.
- Quantitative analysis of 4 × 4 Mueller matrix transformation parameters for biomedical imaging. Photonics, 6(1):34, 2019.
- Polarimetric images of biological tissues based on the arrow decomposition of Mueller matrices. Photonics, 10(6):669, 2023.
- Mueller matrix decomposition for polarized light assessment of biological tissues. J Biophotonics, 2(3):145–56, 2009.
- J. S. Tyo. Design of optimal polarimeters: maximization of signal-to-noise ratio and minimization of systematic error. Appl. Opt., 41(4):619–630, 2002.
- General methods for optimized design and calibration of Mueller polarimeters. Thin Solid Films, 455:112–119, 2004.
- N. C. Bruce and I. Montes-González. The effect of experimental errors on optimized polarimeters. Proc. SPIE, Polarization Science and Remote Sensing XI, 12690:141–147, 2023.
- General and self-consistent method for the calibration of polarization modulators, polarimeters, and Mueller-matrix ellipsometers. Appl. Opt., 38:3490–3502, 1999.
- S. R. Cloude. Conditions for the physical realisability of matrix operators in polarimetry. Proc. SPIE, 116:177–185, 1989.
- The Feynman Lectures on Physics. volume III, chapter 11. Pearson/Addison-Wesley, 2006.
- In vivo imaging of uterine cervix with a Mueller polarimetric colposcope. Sci. Rep., 7:2471, 2017.
- J. Qi and D. S. Elson. Mueller polarimetric imaging for surgical and diagnostic applications: a review. J. Biophotonics, 10(8):950–982, 2017.
- T. Novikova and J. C. Ramella-Roman. Is a complete Mueller matrix necessary in biomedical imaging? Opt. Lett., 47(21):5549–5552, 2022.
- J. J. Gil. On optimal filtering of measured Mueller matrices. Appl. Opt., 55(20):5449–5455, 2016.
- V. Tuchin. Tissue Optics : Light Scattering Methods and Instruments for Medical Diagnosis. (2nd Ed. SPIE Press), 2007.
- Are indices of polarimetric purity excellent metrics for object identification in scattering media? Remote Sens., 14(17):4148, 2022.
- Characterization of the Mueller matrix: Purity space and reflectance imaging. Photonics, 9(2):88, 2022.
- C. N. Haddad. Cholesky factorization. In Encyclopedia of Optimization, pages 374–377. C. A. Floudas, P. M. Pardalos, (Eds) Springer US, Boston, MA, 2009.
- J. J. Gil. Characteristic properties of Mueller matrices. J. Opt. Soc. Am. A, 17(2):328–334, 2000.
- G. T. Gilbert. Positive definite matrices and Sylvester’s criterion. American Math. Mon., 98(1):44–46, 1991.
- B. P. Brooks. The coefficients of the characteristic polynomial in terms of the eigenvalues and the elements of an n×n𝑛𝑛n\times nitalic_n × italic_n matrix. Appl. Math. Lett., 19(6):511–515, 2006.
- X. Wang. A simple proof of Descartes’s rule of signs. The American Mathematical Monthly, 111(6):525–525, 2004.
- I. G. Macdonald. Symmetric functions and Hall polynomials. Oxford Science Publications, 1995.
- Eigen v3. http://eigen.tuxfamily.org, 2010.