The design of a time-interleaved analog-digital conversion modulator based on FPGA-TDC for PET application (2201.01992v1)

Published 6 Jan 2022 in physics.ins-det

Abstract: Fully Field Programmable Gate Array (FPGA)based digitizer for high-resolution time and energy measurement is an attractive low cost solution for the readout electronics in positron emission computed tomography (PET)detector. In recent years, the FPGA based time-digital converter (FPGA-TDC) has been widely used for time measurement in the commercial PET scanners. Yet, for the energy measurement, few studies have been reported on a fully FPGA based, large dynamic range and high resolution alternative to the commercial analog-digital converter (ADC). Our previous research presents a 25 Ms/s FPGA-TDC based free-running ADC (FPGA-ADC), and successfully employed it in the readout electronics for PET detector. In this work-in-progress study, by means of the time-interleaved strategy, a 50 Ms/s FPGA-ADC is presented. With only two off-chip resistors, both the A/D conversion and energy measurement are achieved on a Xilinx Kintex-7 FPGA. Therefore, this method has great advantages inimproving system integration. Initial performance tests are also presented, and we hope it can give us a possibility to develop a new FPGA-only front-end digitizer for PET in future.